2-4
Automated Verification
Step 2. Measurement Calibration
3Insert the verification kit disk into the Agilent 8702Ds disk drive.
4Press PRESET, SAVE/RECALL, SELECT DISK, INTERNAL DISK.
5If you want a printout of the verification data for all the devices, press SYSTEM,
SERVICE MENU, TEST OPTIONS, RECORD ON.
If you switch on the record function, you cannot switch it off during the verifi-
cation procedure.
6Position the paper in the printer so that printing starts at the top of the page.
7Press SYSTEM, SERVICE MENU, TESTS, SYS VER TESTS, EXECUTE TEST.
8When the Agilent8702D displays, Sys Ver Init DONE, the initialization
procedure is complete.
Do not press PRESET or recall another instrument state. You must use the in-
strument state that you loaded during the initialization procedure.
Step 2. Measurement Calibration
1Press CAL, CAL KIT & STDS, SELECT CAL KIT, CAL KIT:7mm, RETURN, RETURN, CALIBRATE
MENU, FULL 2-PORT.
2Press ISOLATION, OMIT ISOLATION, ISOLATION DONE.
3Press REFLECTION.
4Connect the open end of the open/short combination (supplied in the
calibration kit) to reference test port 1, as shown in Figure2-2.