3-4
Manual Verification
Step 2. Measurement Calibration
CAUTION Do not press PRESET or recall another instrument state. You must use the
instrument state that you loaded during the initialization procedure.
Step 2. Measurement Calibration
1Press MENU, POWER, 2, x1 to set the Agilent 8702D test port power to 2 dBm.
2Press CAL, CAL KITS & STDS, SELECT CAL KIT, CAL KIT:7mm, RETURN, RETURN, CALIBRATE
MENU, FULL 2-PORT.
3Press ISOLATION, OMIT ISOLATION, ISOLATION DONE.
4Press REFLECTION.
5Connect the open end of the open/short combination (supplied in the
calibration kit) to reference test port 1, as shown in Figure3-1.
6Press FORWARD:OPEN.
7When the Agilent 8702D finishes measuring the standard, connect the short
end of the open/short combination to reference test port 1.
8Press FORWARD:SHORT.
9When the Agilent8702D finishes measuring the standard, connect the 50 ohm
termination (supplied in the calibration kit) to reference test port 1.
10 Press REFLECTION, FORWARD:LOAD.
11 When the Agilent8702D finishes measuring the standard, connect the open
end of the open/short combination to reference test port 2, as shown in
Figure3-1.
12 Press REVERSE:OPEN.
13 When the Agilent 8702D finishes measuring the standard, connect the short
end of the open/short calibration standard to reference test port 2.
14 Press REVERSE:SHORT.
15 When the Agilent8702D finishes measuring the standard, connect the 50 ohm
termination to reference test port 2.
16 Press REVERSE:LOAD.
17 When the Agilent8702D finishes measuring the standard, press STANDARDS
DONE.