6-65
Performance Tests Option 011
16. Source and Receiver Harmonics (Option 002 Only)
24 Move the receiver A input cable to receiver input B.
25 Press SYSTEM, HARMONIC MEAS, HARMONIC OFF, STOP, 1.5, G/n.
26 Only for Analyzers with Option 006: Press STOP, 3, G/n.
27 Repeat Step 6 through Step 9 to measure the source and receiver worst case
2nd harmonic for the receiver input B. Press MEAS, INPUT PORTS, B in Step 6.
28 Repeat Step 11 through Step 16 to measure the source and receiver worst case
3nd harmonic for the receiver input B.
Receiver Harmonics
29 Connect the equipment as shown in Figure 6-34.
Figure6-34. Receiver Harmonics Test Setup
30 Press SYSTEM, HARMONIC MEAS, HARMONIC OFF, STOP, 1.5, G/n.
31 Only for Analyzers with Option 006: Press STOP, 3, G/n.
32 Repeat Step 6 through Step 9 to measure the receiver worst case 2nd harmonic
for the receiver input B. Press MEAS, INPUT PORTS, B in Step 6.
33 Repeat Step 11 through Step 16 to measure the receiver worst case 3nd
harmonic for the receiver input B.
34 Press SYSTEM, HARMONIC MEAS, HARMONIC OFF.
35 Repeat Step 6 through Step 9 to measure the receiver worst case 2nd harmonic