Test 1. Test Port Output Frequency Range and Accuracy

Test Frequencies
(MHz)
Min.
(MHz)
Results Measured
(MHz)
Max.
(MHz)
Measurement
Uncertainty (MHz)
0.03 0.029 999 7 ____________________ 0.030 000 3 ± 0.000 000 036
0.3 0.299 997 ____________________ 0.300 003 ± 0.000 000 360
5.0 4.999 950 ____________________ 5.000 050 ± 0.000 006
16.0 15.999 840 ___________________ _ 16.000 160 ± 0.000 019
31.0 30.999 690 ___________________ _ 31.000 310 ± 0.000 037
60.999 999 60.999 390 _____ _______________ 61.000 610 ± 0.000 073
121.0 120.998 790 ________ ____________ 121.001 210 ± 0.000 145
180.0 179.998 200 ________ ____________ 180.001 800 ± 0.000 216
310.0 309.995 900 ________ ____________ 310.003 100 ± 0.000 372
700.0 699.930 000 ________ ____________ 700.007 000 ± 0.000 840
1 300.0 1 299.987 ____________________ 1 300.013 ± 0.0 01 560
2 000.0 1 999.980 ____________________ 2 000.020 ± 0.0 02 400
3 000.0 2 999.970 ____________________ 3 000.030 ± 0.0 03 600
Option 006
4.0 3.999 960 ____________________ 4.000 040 ± 0.004 800
5.0 4.999 950 ____________________ 5.000 050 ± 0.006 000
6.0 5.999 940 ____________________ 6.000 060 ± 0.007 200