4-70
Performance Tests
18. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 Only)
harmonic.
46 Press SYSTEM, HARMONIC MEAS, HARMONIC OFF.
47 After one sweep, press DISPLAY, DATAMEMORY, DATA/MEM to normalize the
trace.
48 Press SCALE REF, AUTO SCALE, SCALE/DIV, 1, x1 to get a better viewing of the trace.
49 Press SYSTEM, HARMONIC MEAS, HARMONIC THIRD.
50 After one sweep, press SCALE REF, AUTO SCALE.
51 Press MARKER FCTN, MARKER MODE MENU, MKR SEARCH, SEARCH: MAX.
52 Write the marker 1 value on the Performance Test Record.