09.95
Siemens AG 2001 All Rights Reserved 6FC5197jAA50 4–55
SINUMERIK 840C (IA)
Activating the hard The hard disk is configured with 5 Mbytes for user data. It is possible to use
disk options more memory area if one or several hard disk options are activated. A hard disk
(up to SW 3 only) option makes one of three areas of the hard disk available for additional user
data: an area of 5 Mbytes, an area of 10 Mbytes and an area of 20 Mbytes.
How to activate hard First press the softkey PC START-UP/BACKUP in MMC area DIAGNOSIS under
disk options menu item Setup/Add Options; Add user memory options.
Passwords The following passwords have been allocated:
Menu item 1 MEMO1
2 MEMO2
3 MEMO3
When upgrading to software version >2, these options must first be enabled.
4.7 Customer UMS
General remarks The customer UMS is created using the SW 800A workstation.
The UMS data are given the identifier %UMS and assigned to a language direc-
tory (e.g. ENGLISH) in the SERVICES area. Several UMS files with different na-
mes can exist side-by-side in the language directory. On POWER OFF and ON,
only the file with the name UMS can be loaded from the current language direc-
tory (set language) to the NCK.
An exact description of the operating sequences for loading data from an external
data storage device is to be found in the Operators Guide.
Up to SW 3 Up to SW 3 the customer UMS is enabled with an entry (ASMLEN) in the CON-
FIG file under Master control.
As from SW 4 With SW 4 and higher, please follow the description Flexible memory configura-
tion, see Section Functional Descriptions”.
Note The standard UMS is contained in the scope of supply of the SINUMERIK 840C.
If the user configures his own UMS, the standard UMS is not loaded in the NCK.
Menu 48, The system setting may not allow you to leave menu 48 in the programming area
RECALL Key using the RECALL key. If this happens, the setting in the BEDCONF file must be
changed. See the entry BAPPLMENMODLIST in the following figure:
23 i 2 2 15000 ’0’ ’0’ ’0’ ’0’ ’0’ ’0’ ’0’ ’0’ ’0’ ’134’ ’48’ // BAPPLMENINDLIST
24 i 1 1 15000 // BAPPLIND
25 i 2 2 15001 ’0’ ’0’ ’0’ ’0’ ’0’ ’0’ ’0’ ’0’ ’0’ ’0’ ’4’ // BAPPLMENMODLIST
26 i 1 1 15000 // BAPPLIND
27 i 2 2 15002 ’0’ ’1’ ’0’ ’0’ ’0’ ’0’ ’0’ ’0’ ’2’ ’1’ ’1’ // BAPPLNCBERLIST
4 MMC Area Diagnosis
4.7 Customer UMS