Manual Calibration

Applying new constants to the data file

NOTE: If you are calibrating Mariner data, see “Ensuring that masses match during calibration” on page 5-24.

To apply the calibration constants to the data:

If you are

Click

The following occurs

calibrating

 

 

 

 

 

Mariner data

Apply

All spectra in the data file are calibrated and

(including

Calibration

displayed with an MC trace label. The calibration

MS Method

 

constants are saved with the data file. Each

data)

 

spectrum in the data file is calibrated when

 

 

displayed.

 

 

 

Voyager data

Apply

The current spectrum is calibrated and

 

Calibration

displayed with an MC trace label. The calibration

 

 

constants are saved with the spectrum.

 

 

 

 

Apply to All

All spectra in the data file are calibrated using

 

NOTE: This

the currently displayed calibration, and are

 

button is

displayed with an MC trace label. The calibration

 

constants are saved with the data file. Each

 

displayed only if

 

spectrum in the data file is calibrated when

 

you are

 

displayed.

 

calibrating a

 

 

 

Voyager

NOTE: If you select Apply to All, you overwrite

 

multispectrum

any calibrations previously applied to individual

 

data file.

spectra by clicking Apply Calibration.

 

 

 

5

Data ExplorerSoftware User’s Guide

5-15