Chapter 5 Examining Spectrum Data
Refer to the following table to determine how to set Advance Baseline Correction parameters and obtain the desired baseline correction.
Condition |
| Set Parameters To |
|
|
|
Baseline follows a rising peak cluster | • | Peak Width— Average peak width |
| • | Flexibility— 0.5 |
| • | Degree— 0.5 |
Baseline is gently curving | • | Peak Width— 10 to 20 times peak width |
| • | Flexibility— 0 |
| • Degree— 0.5 (decrease further to raise | |
|
| the baseline) |
Baseline rise should be ignored and | • | Peak Width— Half of the width of the | ||
treated as signal |
| cluster to correct (in the illustration | ||
| Peak cluster |
| shown, if the width of the cluster is 200, | |
|
| set Peak Width to 100) | ||
|
|
| • | Flexibility— 0 |
|
|
| ||
|
|
| • Degree— 0.5 (decrease further to raise | |
|
|
| ||
|
|
|
| the baseline) |
5