Peak Detection and Labeling Troubleshooting

Table 9-12 Charge State and Isotope Determination Troubleshooting—Mariner

Only (Continued)

Symptom

Possible Cause

Action

 

 

 

 

 

 

 

Known isotope not labeled

Charge State peak labels

Turn on Charge State

 

 

with charge state

disabled

peak labels. See

 

 

 

 

Section 3.5.2, Setting

 

 

 

 

Chromatogram

 

 

 

 

and Spectrum Peak

 

 

 

 

Labels.

 

 

 

 

 

 

 

 

Max Charge State

Set Max Charge State

 

 

 

parameter set too low.

correctly. See “Peak

 

 

 

See example in “Max

Processing parameters

 

 

 

Charge State set too low”

(spectrum data only)” on

 

 

 

on page 3-33.

page 3-26.

 

 

 

 

 

 

 

 

Peak List Filtering is

Disable Peak List

 

 

 

enabled, with Charge

Filtering. See “Filtering

 

 

 

State filter enabled and

the spectrum peak list” on

 

 

 

set too low

page 3-42.

 

 

 

 

 

 

 

 

Minimum Intensity set too

Set Minimum Intensity

 

 

 

 

 

high to detect other

correctly. See “Advanced

 

 

 

isotope peaks. See

Settings (spectrum data

 

 

 

example in “Effect of

only)” on page 3-28.

 

9

 

Minimum Intensity” on

 

 

 

page 3-36.

 

 

 

 

 

 

 

 

 

Filter width set too high to

Set Filter Width correctly.

 

 

 

determine other isotope

See “Advanced Settings

 

 

 

peaks

(spectrum data only)” on

 

 

 

 

page 3-28.

 

 

 

 

 

 

 

Data ExplorerSoftware User’s Guide

9-19