Chapter 5 Examining Spectrum Data

5.3.4 Reverting to Instrument Calibration

The Revert to Instrument Calibration function does the following:

Mariner data— Reapplies the original calibration constants used to acquire the data.

Voyager data— Applies default calibration to the data, regardless of whether default or external calibration is used to acquire the data.

To revert the calibration:

1.From the Process menu, select Mass Calibration, then select Revert to Instrument Calibration.

The spectrum is recalibrated with the original calibration constants used during acquisition (Mariner data) or default calibration (Voyager data) and displayed with an MC trace label. The original and new constants are displayed in the Output window.

2.To save the reverted calibration with the data file, select Mass Calibration from the Process menu, then:

If you are

Select

The following occurs

calibrating

 

 

 

 

 

Mariner

Apply Calibration

All spectra in the data file are calibrated and

data

 

displayed with an MC trace label. The

(including

 

calibration constants are saved with the data

MS Method

 

file. Each spectrum in the data file is

data)

 

calibrated when displayed.

 

 

 