Chapter 5 Examining Spectrum Data
6.Specify reference masses to add by doing either of the following:
•Click a mass, then click OK.
•Type new reference mass information in the Name, Theoretical m/z, Charge, and Elemental Composition fields, then click OK.
NOTE: You must type a minus sign
NOTE: You can save changes to the Calibration Reference File by clicking Save. For more information, see “Modifying a calibration reference file” on page
The Automatic Calibration Settings dialog box is displayed again (see Figure
7. | ||
8. | To delete masses, do either of the following: | |
|
| • Select a reference, then click Delete Selected |
|
| Reference |
|
| • Click Delete All References |
9. | Enter Reference Matching Criteria: | |
|
| • Minimum Intensity— Peaks must be above this |
|
| intensity to be considered a match. Select the unit |
5 |
| for Minimum Intensity, % Relative Intensity or |
| Absolute Counts. | |
|
| |
|
| • Mass Tolerance— Peaks must be within this |
|
| tolerance of the theoretical m/z to be considered a |
|
| match. Select the unit, m/z or ppm. |