Peak Detection and Labeling Troubleshooting

Table 9-10 Peak Detection and Labeling Troubleshooting—Mariner and Voyager

(Continued)

Symptom

Possible Cause

Action

 

 

 

Expected user label not

Delta X value includes

Set Delta X value low

displayed

more than one peak apex

enough to prevent the

 

 

peak labeling windows

 

 

from overlapping. For

 

 

information, see

 

 

Section 3.5.3, Setting

 

 

Custom Peak Labels.

 

 

 

 

Delta X value is outside

Set Delta X value within

 

acquisition range for the

acquisition range for the

 

data file

data file. For information,

 

 

see Section 3.5.3, Setting

 

 

Custom Peak Labels.

 

 

 

Peak label placed on peak

Filter Width Increment set

Set Increment to 1. See

shoulder instead of peak

higher than 1

“Increment” on page 3-31.

apex

 

 

 

 

 

Table 9-11 Peak Detection and Labeling Troubleshooting—Voyager Only

9

Symptom

Possible Cause

Action

 

 

 

Noise detected as peaks

%Max Peak Area set too

Increase. See

 

low

Section 3.2.2, Strategy for

 

 

Voyager Peak Detection.

 

 

 

Low Mass Gate spike

Software identifies the

Truncate the data. See

identified as Base Peak

most intense peak in a

Section 5.9, Truncating a

(linear data)

trace as the base peak,

Spectrum.

 

and does not ignore

 

 

artifacts

 

 

 

 

Data ExplorerSoftware User’s Guide