Chapter 9 Troubleshooting

Table 9-7 Calibration Troubleshooting—Mariner Only

Symptom

Possible Cause

Action

 

 

 

Mass Calibration

Chromatogram window is

Select Spectrum window.

commands are dimmed

selected

 

 

 

 

Apply Calibration

For MS Method data,

No action. Normal

command is dimmed

calibration is valid for an

occurrence.

when calibrating

individual spectrum. You

 

MS Method data

cannot apply the

 

 

calibration from one

 

 

spectrum to the entire

 

 

data file.

 

 

 

 

Error displayed when you

.CAL file corrupted

Create new .CAL file. See

import a calibration

 

“Exporting .BIC, .MSM,

 

 

and .CAL files” on

 

 

page 1-36.

 

 

 

 

Importing a .CAL file

Import a Mariner .CAL.

 

generated from a Voyager

 

 

data file

 

 

 

 

9

9-12 Applied Biosystems