Chapter 9 Troubleshooting

Table 9-12 Charge State and Isotope Determination Troubleshooting—Mariner

Only (Continued)

 

 

Symptom

Possible Cause

Action

 

 

 

 

 

 

 

Spectrum peaks not

Mass of original molecule

No action. Normal

 

 

labeled with charge state

above 4,000 Da, not

occurrence.

 

 

when charge state labels

range in which the Mariner

 

 

 

are selected

system can resolve the

 

 

 

 

isotope peaks

 

 

 

 

 

 

 

 

 

Charge State parameters

Adjust parameters. See

 

 

 

not set to detect at least

“Peak Processing

 

 

 

two isotope peaks

parameters (spectrum

 

 

 

 

data only)” on page 3-26.

 

 

 

 

 

 

 

 

Peaks are more than 1 Da

No action. Normal

 

 

 

apart

occurrence.

 

 

 

 

 

 

 

 

Peaks are not from the

No action. Normal

 

 

 

same isotope species

occurrence.

 

 

 

 

 

 

 

 

Filter width is set too high

Adjust Filter Width. See

 

 

 

to detect other isotope

“Advanced Settings

 

 

 

peaks

(spectrum data only)” on

 

 

 

 

page 3-28.

9

 

 

 

 

 

 

Maximum Charge State

Adjust the Maximum

 

 

for charge state

Charge State. See “Peak

 

 

 

determination is set lower

Processing parameters

 

 

 

than the charge state of

(spectrum data only)” on

 

 

 

the peak

page 3-26.

 

 

 

 

 

9-20 Applied Biosystems