Chapter 8 Viewing Voyager PSD Data

Solving and plotting

Applying new constants to the data file

8

After matching peaks, click Solve and Plot.

The calibration statistics are displayed in the Result tab of the Output window and the calibration constants are applied to the spectrum displayed

If you calibrate more than one time, subsequent calibration statistics are added to the end of the list in the Output window. Older calibration statistics are listed at the top of the list. Use the scroll bar to view newer statistics at the bottom of the list.

The spectrum is calibrated and displayed with an MC trace label. The calibration statistics are displayed in the Output window.

To apply the calibration constants to each spectrum (all segments) in the data file, click Apply Calibration.

All spectra in the data file are calibrated and displayed with an MC trace label. The calibration constants are saved with the data file. Each spectrum in the data file is calibrated when displayed.

The software uses different values in the calibration equation calculation based on the type of calibration performed.

8-18 Applied Biosystems