Processing, Tools, and Applications Troubleshooting

Table 9-5 Processing, Tools, and Applications Troubleshooting—Mariner Only

Symptom

Possible Cause

 

Action

 

 

 

 

 

 

 

Failed to calculate result

Did not select at least two

Select at least two peaks

 

 

for mass deconvolution

peaks for same charge

(for example, +1 and +2).

 

 

 

envelope

For more information, see

 

 

 

 

Section 5.6, Mass

 

 

 

 

Deconvolution

 

 

 

 

(Mariner Data Only).

 

 

 

 

 

 

 

Multiple Charge/Mass

Your system does not

Contact Applied

 

 

Deconvolution commands

include the optional mass

Biosystems to purchase

 

 

dimmed on Process menu

deconvolution software

the option.

 

 

 

 

 

 

 

Centroiding, Mass

Spectrum window not

Activate Spectrum

 

 

Calibration, Multiple

active

window, then select

 

 

Charge commands not

 

Process menu.

 

 

displayed on Process

 

 

 

 

 

menu

 

 

 

 

 

 

 

 

 

 

Resolution command not

Spectrum window not

Activate Spectrum

 

 

 

displayed on Tools menu

active

window, then select Tools

 

 

 

 

menu.

 

 

 

 

 

 

 

 

9

Only the active trace

Zooming may not behave

Click

in the top right

 

 

 

zooms in Overlay mode

as expected on overlaid

corner of the window to

 

 

 

traces

restore the view.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Data ExplorerSoftware User’s Guide