Calibrating a PSD Spectrum

Matching peaks automatically

If you want the software to compare observed masses in all segment spectra included in the .DAT file to reference masses in the selected calibration reference file:

1.Click Match.

If a mass within the tolerance of any of the masses listed in the calibration reference file is found in any spectrum in the .DAT file, the match is displayed in the Calibration Mass Peak Selection window.

CAUTION

Use the Match function with care. Before using, ensure

 

that all peaks in all segments are correctly peak

 

detected. The Match function examines the complete

 

mass range in all segments in the .DAT file.

 

8

 

 

 

 

NOTE: The entire mass range of each spectrum in the

.DAT file is checked, not just the mass range included

in the composite spectrum, and even if the spectrum is

 

not currently displayed in the Spectrum window.

 

 

 

 

For comparison, the difference between the reference mass in the calibration reference file and the observed peak mass is displayed.

Hint: You can sort the list of matches by clicking on the column header buttons (see Figure 8-5 on page 8-13). You can display complete information about a reference mass by double-clicking on the mass.

If no matches are found, an error message is displayed.

2.To delete a match from the list, select the mass and click Delete Selected Match.

To clear the entire list, click Delete Entire List.

3.Complete the calibration as described in “Solving and plotting” on page 8-18.

Data ExplorerSoftware User’s Guide

8-15