Scale Error Reporting

Figure 4-1. 7-Segment LED Display

7-Segment Display

Scale Error Reporting

Scale diagnostics uses the Remote Scale Display and the Zero Status lamp to communicate specific scale failures. The following chart shows the Remote Display messages, the Scale Status lamp indication, the problem that the scale is experiencing and what action should be taken. When troubleshooting, always remember to check all cable connections first before proceeding with other problem isolation steps.

Remote

Scale Status

Problem

Display

Lamp

Description

Action Required

Flash, pause, 1 blink,

E _1 long pause, repeat sequence.

Flash, pause, 2

E__2 blinks, long pause, repeat sequence.

Too much motion at power- up.

Calibration lost.

Check for stable installation. Change scale motion filter using the labels found in Chapter 6; restart. If prob- lem persists, scale may require calibra- tion.

Call technical support personnel. Re- calibrate; restart; recertify (if required).