8. The Yellow LED will extinguish for approximately 10 seconds and
the Remote Display will alternately display and until the scale is ready to proceed.
9.If the calibration was successful, the speaker sounds a single tone, the Scale Status LED begins blinking again, and
10.If the calibration was not successful, the speaker will sound five tones indicating a scale failure, and the Scale Status LED will blink twice, strobe (fast blinks) and then continually repeat this sequence until reset. Remove all weight from the Weigh Platter, and repeat the procedure starting with step 7.
11.Press the Zero Push Button again to permanently store the calibra- tion data and exit Calibration Mode. You have completed the cali- bration of the scanner/scale. The unit’s calibration must now be verified as required by state and/or local weights and measures regu- lations. The verification procedure must be performed to assure that a scale will pass Weights and Measures requirements before it is placed into commercial/retail service.
You have completed the scale calibration procedure. You must now con- tinue with the calibration verification tests to complete the scale’s calibra- tion. If you are weighing in pounds, continue with the following procedures to verify that the scale was successfully calibrated and that it weighs properly. If you are weighing in kilograms, turn to the instructions titled, Calibration Verification (Kilograms), later in this chapter.
Magellan® 9500 |