Calibration Verification (Kilograms)

Calibration Verification (Kilograms)

Once you have completed the calibration sequence, you may be required to perform these step-by-step verification procedures. These procedures follow the National Institute of Standards and Technology Handbook-44guidelines for grocery scale installations. You may be required by state or local law to have these procedures performed by a certified technician or verified by a proper official.

These calibration verification procedures cover five different tests:

Increasing-Load Test

Shift Test

Blanking Test

Decreasing-Load Test

Return to Zero Test

Increasing-Load Test (Phase 1)

This test checks Scale operation for increasing loads from 0.100 kg and 7.50 kg.

1.Check that the display reads 0.000 kg when at rest with nothing on the weighing surface. (The Yellow LED is steadily lit).

2.Place a 100 gram weight on the center of the weighing surface and check that the display reads 0.100 kg.

3.Place an additional 200 grams on the center of the weighing surface and check that the display reads 0.300 kg.

4.Place an additional 200 grams on the center of the weighing surface and check that the display reads 0.500 kg.

5.Place an additional 100 grams on the center of the weighing surface and check that the display reads 0.600 kg.

6.Place an additional 100 grams on the center of the weighing surface and check that the display reads 0.700 kg.

7.Place an additional 100 grams on the center of the weighing surface and check that the display reads 0.800 kg.

Product Reference Guide