Note

Another feature of markers set to patterns is the Stop measurement when X-O ___ which is found in the Specify Patterns field. The options are:

Off

Less than

Greater than

In range

Not in range

This feature is only available when Count is set to Time in the Trace menu. With this feature you can use the logic analyzer to look for a specified time or range of time between the marked patterns and to stop acquiring data when it finds this time between markers. The X marker must precede the O marker.

The upper and lower range boundaries must not be the same value. For example, if you want to stop a measurement when the X and O markers are in range of 200 ns, you should set the range values to 190 ns and 210 ns. This eliminates erroneous measurement termination.

Markers Time When the markers are set to Time, you can place the markers on states of interest in the listing and the logic analyzer will tell you:

Trig(ger) to X

Trig(ger) to O

Time X to O

To position the markers, touch the field of the marker you wish to position.

Figure 6-14. Markers Time

Interpreting the Display

HP 16510B

6 - 14

Front-Panel Reference

Page 142
Image 142
HP 16500A manual Markers Time Interpreting the Display HP 16510B