Probe Cable Installation
Installation
The HP 16510B State/Timing Module comes with probe cables installed by the factory. If a cable is to be switched or replaced, refer to "Probe Cable Replacement" in Section VI of the HP 16510B Service Manual.
Caution
Module
Installation
Caution
Installation Considerations
Do not install, remove or replace the module in the instrument unless the instrument power is turned off.
The HP 16510B State/Timing Module will take up one slot in the card cage. For every additional HP 16510B State/Timing Module you install, you will need an additional slot. They may be installed in any slot and in any order. Procedures for installing the logic analyzer module cards are shown in the
The following procedure is for the installation of the HP 16510B Logic Analyzer Module.
The effects of ELECTROSTATIC DISCHARGE can damage electronic components. Use grounded wrist straps and mats when you are performing any kind of service to this module.
∙The HP 16510B State/Timing Module(s) can be installed in any available card slot and in any order.
∙Cards or filler panels below the empty slots intended for module installation do not have to be removed.
∙The probe cables do not have to be removed to install the module.
Installing New Logic Analyzer Boards into the Mainframe | HP 16510B |
|