What Am I Going to Measure?

After configuring the logic analyzer and hooking it up to your circuit under test, you will be measuring the time (x) from when the RAS goes low to when the CAS goes high, as shown below.

Figure 7-1. RAS and CAS Signals

How Do I

Configure the

Logic Analyzer?

In order to make this timing measurement, you must configure the logic analyzer as a timing analyzer. By following these steps you will configure Analyzer 1 as the timing analyzer.

If you are in the State/Timing E Configuration menu you are in the right place and you can start with step 2; otherwise, start with step 1.

1.Touch the field in the upper left corner of the display and select State/Timing E.

2.In the State/Timing E Configuration menu, change Analyzer 1 type to Timing. If Analyzer 1 is already a timing analyzer, go on to step 3.

a.Touch the field Type: _________ .

b.When the pop-up appears, touch Timing.

Using the Timing Analyzer

HP 16510B

7 - 2