Symbols Pattern Symbols: User can define a mnemonic for the specific bit pattern of a label. When data display is SYMBOL, mnemonic is displayed where the bit pattern occurs. Bit pattern can include 0s, 1s, and don’t cares.
Range Symbols: User can define a mnemonic covering a range of values. Bit pattern for lower and upper limits must be defined as a pattern of 0s and 1s. When data display is SYMBOL, values within the specified range are displayed as mnemonic ± offset from base of range.
Number of Pattern and Range Symbols: 100 per analyzer. Symbols can be
Timing Analysis
Transitional Timing Mode
Sample is stored in acquisition memory only when the data changes. A time tag stored with each sample allows reconstruction of waveform display. Time covered by a full memory acquisition varies with the number of pattern changes in the data.
Sample Period: 10 ns.
Maximum Time Covered By Data: 5000 seconds.
Minimum Time Covered By Data: 10.24 μs.
Glitch Capture Mode
Data sample and glitch information stored every sample period.
Sample Period: 20 ns to 50 ms in a
Memory Depth: 512 samples/channel.
Time Covered by Data: Sample period × 512.
HP 16510B | Specifications and Operating Characteristics |