Measurement

and Display

Functions

Autoscale (Timing Analyzer Only)

Acquisition Specifications

Autoscale searches for and displays channels with activity on the pods assigned to the timing analyzer.

Arming: Each analyzer can be armed by the run key, the other analyzer, or the Intermodule Bus.

Trace Mode: Single mode acquires data once per trace specification; repetitive mode repeats single mode acquisitions until stop is pressed or until time interval between two specified patterns is less than or greater than a specified value, or within or not within a specified range. There is only one trace mode when two analyzers are on.

Labels Channels may be grouped together and given a six character name. Up to 20 labels in each analyzer may be assigned with up to 32 channels per label. Primary use is for naming groups of channels such as address, data, and control busses.

Indicators Activity Indicators: Provided in the Configuration, State Format, and Timing Format menus for identifying high, low, or changing states on the inputs.

Markers: Two markers (X and 0) are shown as dashed lines on the display.

Trigger: Displayed as a vertical dashed line in the timing waveform display and as line 0 in the state listing display.

HP 16510B

Specifications and Operating Characteristics

Front-Panel Reference

C-7

Page 261
Image 261
HP 16510B, 16500A manual Measurement Display Functions, Autoscale Timing Analyzer Only Acquisition Specifications