Probe Cable
Caution
The probe pod cable contains 17 signal lines, 34 chassis ground lines and two power lines that is woven together. It is 4.5 feet long.
The probe grounds are chassis (earth) grounds, not "floating" grounds.
Each cable is capable of carrying 0.67 amps for preprocessor power. Current in excess of 0.67 amps per cable will cause the preprocessor supply voltage to drop below a safe level. DO NOT exceed this 0.67 amps per cable or the preprocessor may malfunction. Also, the maximum power available from the logic analyzer (all cables) is 2 amps at 5 volts.
Note | The preprocessor power source is fused. The fuse is located inside the HP | ||||
|
|
|
|
| 16500A on the logic analyzer card. If a preprocessor appears to be |
|
|
|
|
| |
|
|
|
|
| malfunctioning, refer to the HP 16510B service manual for instructions on |
|
|
|
|
| |
|
|
|
|
| checking this fuse. |
|
|
|
|
|
|
|
|
|
|
| The probe cable connects the logic analyzer to the pods, termination |
|
|
|
|
| adapter, or the HP 10269C General Purpose Probe Interface. |
Probes Each probe is a
Figure 2-4. Probe Cable
HP 16510B | Probing |
| 2 - 5 |