Measurement Configurations
Analyzer Configurations:
Analyzer 1 | Analyzer 2 |
Timing | Off |
Off | Timing |
State | Off |
Off | State |
Timing | State |
State | Timing |
State | State |
Off | Off |
Channel Assignment: Each group of 16 channels (a pod) can be assigned to Analyzer 1, Analyzer 2, or remain unassigned. The HP 16510B contains 5 pods.
State Analysis
Memory Data Acquisition: 1024 samples/channel.
Trace Clocks: Five clocks are available and can be used by either one or two
Specification state analyzers at any time. Clock edges can be ORed together and operate in single phase, two phase demultiplexing, or two phase mixed mode.
Clock edge is selectable as positive, negative, or both edges for each clock.
Clock Qualifier: The high or low level of up to four clocks can be
ANDed with the clock specification. Setup time: 20 ns; hold time: 5 ns.
Pattern Recognizers: Each recognizer is the AND combination of bit (0, 1, or X) patterns in each label. Eight pattern recognizers are available when one state analyzer is on. Four are available to each analyzer when two state analyzers are on.
HP 16510B | Specifications and Operating Characteristics |