Philips V24CT, V26CT manual Parameter Settings Transfer Messages, Message Condition Action required

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 137
Image 137

Parameter Settings Transfer Messages

Parameter Settings Transfer Messages

The following messages related to Parameter Settings Transfer can appear on the display:

Message

Condition

Action required

 

 

 

 

 

(Parameter name)

The indicated parameter

Insert a module that has

no settings

module is not a T

a T on the front.

transfer capability

module.

 

 

 

 

 

 

 

 

(Parameter name)

No settings are stored in

Enter the parameter

check settings''

the module, or the

Task Window and check

 

settings are incorrectly

the settings. The

 

stored.

settings are

 

 

automatically stored

 

 

when you exit the Task

 

 

Window.

 

 

 

 

 

Check Patient

The patient category

Either change the

Category and

stored with the settings

patient category for the

(parameter name)

in the module is

monitor or change the

settings

different from that of the

parameter settings.

 

monitor.

 

 

 

 

 

 

 

 

(Parameter name)

The parameter module

Press

 

,

Silence Reset

unable to store

was removed before the

re-insert the module and

settings in module

settings could be stored,

check/adjust the

 

or there is a malfunction

settings in the

 

in the module. (This

parameter Task

 

message can appear with

Window. Exit the Task

 

Settings Transfer On or

Window and leave the

 

Off.)

module in the rack for

 

 

about 10 seconds before

 

 

you remove it. If the

 

 

same message appears,

 

 

replace the module.

 

 

 

 

 

Setting up your Monitor

Setting up your Monitor 3-63

Page 137
Image 137