Maintenance

Performance Assurance Checks

 

 

 

Performing

Auto Check - The check is performed automatically when the system is

the tcpO2/

switched on, during calibration and continuously when no transducer is

tcpCO2

connected. When the auto check fails, the INOP message “Equipment

Module Self-

Malfunction'' is given (refer to the Troubleshooting Chapter in the

Test

service manual for troubleshooting details).

 

Self-Test- The tcpO2/tcpCO2 module self-test may be performed at any

 

time. The test signal lasts about 30 seconds and then the display returns

 

to the normal monitoring mode.

 

 

 

 

TEST - Press

 

 

 

 

 

 

 

Monitor Setup

 

 

 

Test Signals

 

RESULT - If the module is functioning correctly, the following numerics are displayed without the alarms being activated:

tcpO2 60 mmHg (8.0 kPa) tcpCO2 40 mmHg (5.3 kPa)

C aution

Radiated immunity: If operating under conditions according to EN

60601-1-2 up to 1 V/m electrical field strength, the following deviations in the measurements could occur: tcpO2 2 mmHg or less, tcpCO2 1 mmHg or less. It is recommended to avoid the use of electrically radiating devices in close proximity to a tcpO2/tcpCO2 monitor.

13-18 Maintenance