Performance Assurance Checks

Module

Test Numeric

Test Waveform

 

 

 

Cardiac Output

5 l/min (+0.1 l/min)

No wave

 

(Blood Temp: 37°C)

 

 

(Inj Temp: 0°C)

 

 

 

 

tcpO2/tcpCO2

 

No wave

 

 

 

tcpO2

60 mmHg (8.0 kPa)

 

 

 

 

tcpCO2

40 mmHg (5.3 kPa)

 

 

 

 

CO2

 

Simulated square wave

 

 

 

ETCO2

40 mmHg (6.0 kPa)

 

 

 

 

IMCO2

0 mmHg (0.0 kPa)

 

 

 

 

AWRR

25 rpm

 

 

 

 

Temperature

40°C (104°F)

No wave

 

 

 

Maintenance

Performing

the ECG

Module and

ECG/RESP

Self-Test

Auto Check - None

Self-TestThe ECG and ECG/RESP self-test may be performed at any time as the lead selector and pace pulse detection are inhibited during the check. The module will produce ECG test signals at the beginning of the processing circuitry, and process the test signals in the same way patient signals are processed. The test signals last about 30 seconds and then the display returns to the normal monitoring mode.

TEST - Press

Monitor Setup

 

 

 

Test Signals

.

 

 

RESULT - If the module is functioning correctly, the following heart rate and respiration are displayed without the alarms being activated:

ECG 100 bpm in ADULT mode 125 bpm in NEO/PEDI mode

RESP 15 rpm in ADULT mode 30 rpm in PEDI mode 55 rpm in NEO mode

Maintenance 13-15