The calibration pulse varies according to the wave type:

ECG

1 mV square wave

RES

1 ohm M-shaped wave

All Pressures

Stepped wave with 30, 60, 120, and 180 mmHg

 

increments

For pressure scales of 240 and 180, all four steps are displayed. At lower ranges, however, not all the steps are displayed since one or more steps exceeds the full scale range.

Note—If you don't want to insert a calibration pulse or mark an event after you have pressed Mark Event , press any other hardkey (such as Main Screen ).