Philips V24CT, V26CT manual Message, Is displayed if an unknown module is plugged into the rack

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 41
Image 41

Parameter Modules

For most types of parameter modules, the system allows only one of each type per patient (ECG, for example). Other types of modules allow more than one per patient (Invasive Pressure, for example).

If too many modules or an unsupported module are plugged in, a message detailing where the extra module is, appears in the system message field:

Currently ignored module in rack position R-P

where:

Ris the number of the rack

(e.g. 1=integral rack, 2=first satellite rack,...

or 1=first satellite rack, 2=second satellite rack,...

Pis the slot number in that rack (counted from left to right)

The message

Unrecognized module in rack position R-P

is displayed if an unknown module is plugged into the rack.

Note—Since the V24 and V26 Patient Monitors only support one module rack, R will always be 1.

The CMS and V24 and V26 Patient Monitors 1-19

The CMS and V24 and V26 Patient Monitors

Page 41