Philips V26CT, V24CT manual Connectors

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 342
Image 342

Monitor Installation

and Patient Safety

 

 

Controls and Connectors

2.V AMP; this is the vertical amplitude control used to adjust the height of the displayed video.

3.V SHIFT; this is the vertical shift control used to adjust the vertical position of the displayed video.

4.H AMP; this is the horizontal amplitude control used to adjust the width of the displayed video.

5.EAST WEST; this control is used to adjust the pin-cushion effect of the displayed video.

Connectors of

1.

There are two fuse holders for the line protection fuses. The fuses

the M1092A,

 

depend on the type of monitor and the power supply as

M1094A and

 

summarized below:

 

M1094B

 

200-240V M1094A color monitor

2.5A 250V

Displays

 

 

 

200-240V M1094B color monitor

3.15A 250V

 

 

200-240V monochrome monitor

2A 250V

 

 

100-120V color or monochrome monitor

4A 125V

 

 

 

(Slow-Blow)

 

2.

Local Power Connector; a 3-pin connector is used to input the

 

 

local line voltage.

 

 

3.

System Power Connector; a 15-pin “D” type connector is used to

 

 

output 60V DC (100W max) to a similar connector on the DC/DC

 

 

converter in the rear of the computer module.

 

 

4.

Video In Connector; a 9-pin “D” type connector is used to input the

 

 

video signal to be displayed from the display controller in the

 

 

computer module (DSPC for monochrome or CDSPC-Video for

 

 

color).

 

 

5.

Video Out Connector; a 9-pin “D” type connector is used to output

 

 

the video signal to be displayed on a duplicate display.

11-26 Monitor Installation and Patient Safety

Page 342
Image 342
Philips V26CT, V24CT manual Connectors