Philips V24CT, V26CT manual Neonatal Event Review

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 281
Image 281

Viewing Neonatal Events

Event The patient monitors can be adjusted in monitoring mode so that an event Threshold Line is either triggered by an alarm or by criteria defined specifically by the

user.

The event trigger level is represented in the graph by a threshold line, located in each graph channel. This line enables the user to see the following:

1.Whether a trigger is active. This means that the parameter is switched on, the parameter alarm is switched on, the main alarms are switched on and valid numerical values are available. The threshold line is indicated on the graph by an unbroken line.

2.At what value it has been set.

3.Whether it has been changed during the 24-hour review period.

(1)Event Threshold Line

(2)Non Active Threshold Line

(3)Trigger Level Change

Note— When the patient monitor is connected to the Philips Information Center, the Bradycardia threshold line is displayed as a shaded area.

Neonatal Event

Review

 

 

Neonatal Event Review 9-7

Page 281
Image 281
Philips V24CT, V26CT manual Neonatal Event Review