CSA Display (CMS only)

Main Screen

Standard Display

CSA Display

4 realtime waves

CSA + 2 realtime wave

6 realtime waves

CSA + 3 realtime waves

8 realtime waves

CSA + 4 realtime waves

Notes on

• The CSA information is overlayed on the realtime waves at the

CSA

bottom of the Main Screen. These realtime waves disappear from

 

the display, but the parameters are still being monitored, and the

 

parameter alarms are still active.

 

• If CSA covers part of an overlap channel, the entire overlap channel

 

disappears from the display.

 

• If you are using a touchscreen, selecting the CSA display via touch

 

or mouse-click will open up the EEG Setup Task Window