Maintenance Checks

 

 

 

Maintenance

Frequency

Source of Information

 

 

 

 

 

 

Electrical Safety Tests

At least every two years, after

See “Testing and Maintaining the

 

system installation, after any

CMS” in the “CMS Service Guide”

 

repairs, or as needed.

or “Testing and Maintaining the

 

 

V24 and V26“in the “V24 and V26

 

 

Service Guide”.

 

 

 

Testing of all cables and cords.

Daily.

See “Testing and Maintaining the

 

 

CMS” in the “CMS Service Guide”

 

 

or “Testing and Maintaining the

 

 

V24 and V26“in the “V24 and V26

 

 

Service Guide”.

 

 

 

Calibration of the measurement

Every 2 years.

See individual module chapters in

modules where appropriate

 

Volume 2 of this manual for

(For EU/EFTA countries only)

 

details.

 

 

 

User Tests

Module dependent. Please refer to

See chapter 13 “Maintenance” of

 

chapter 13 “Maintenance” of this

this manual for details.

 

manual or chapter 3 of the CMS

 

 

Service Guide “Maintaining the

 

 

CMS” for details.

 

 

 

 

This recommendation does not supersede any local procedures or legal requirements.

Monitor Installation and Patient Safety 11-15