Neonatal Event

Review

 

 

Adjusting Neonatal Event Review Settings

Bradycardia

Alarm Triggered

Event Settings

The Bradycardia event is triggered by a Red*** Brady alarm.

 

 

The Bradycardia alarm limit can be viewed but not adjusted in the Event

 

Setup task window. The alarm limit must be adjusted in Monitoring

 

Mode in the HR/PULSE or Alarm Limits task window and the trigger

 

time must be adjusted in Configuration Mode in the HR/PULSE task

 

window.

 

If the Bradycardia alarm related trigger threshold and trigger time items

 

are selected, the

 

key is inactive and the following

 

Change Content

 

status message is displayed in the Event Setup task window:

 

• “Brady items must be changed in the HR/PULSE Setup”

 

User Defined

 

For situations where the storage and documentation of an event should

 

be independent of all Bradycardia alarms, you can adjust the

 

Bradycardia event threshold and trigger time in Monitoring Mode in

 

the Event Setup task window.

 

Note—When the patient monitor is connected to an Arrhythmia System,

 

the Bradycardia alarms are generated by the latter and not the former.

 

The alarm limits can only be viewed and adjusted on the Arrhythmia. In

 

this situation, the Bradycardia threshold line is displayed in the Event

 

Review task window as a shaded area and the Brady trigger time and

 

threshold items are set to “From Arrhyth” in the Event Setup task

 

window.

9-20 Neonatal Event Review