Philips V26CT, V24CT manual OxyCRG, Alarm Recording

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 210
Image 210
Change AlRecTyp

Recording Functions

oxyCRG Recordings

oxyCRG

Monitor Setup

Alarm

 

Recording

 

Recordng Setup

 

 

 

 

Delayed & Alarm

 

 

 

 

 

Pressing thesoftkey allows you to choose

between “Standard” alarm recording or “oxyCRG” alarm recording.

If you select “oxyCRG”, an oxyCRG recording is started when an alarm occurs. Six minutes of pre-alarm data and two minutes of post-alarm data are recorded. Any manually initiated recording that is in progress will be interrupted, and the oxyCRG recorded instead. If a second alarm occurs while an oxyCRG alarm recording is already in progress, the recording will be extended to two minutes after the second alarm appears on the strip.

If you select “Standard”, your selected realtime waves are recorded when an alarm occurs, as described earlier in this chapter.

6-38 Recording Functions

Page 210
Image 210
Philips V26CT, V24CT manual OxyCRG, Alarm Recording