Making

Calibrated

ECG

Recordings

If the Recorder is Busy

Realtime Wave Recordings

Note—(CMS only): The number of recording channels available depends on the recorder. If you change from a 4- to a 2-channel recorder, only the first two waves are recorded.

Note—Non-preset recordings are made on the same recorder as procedure recordings (selected in the Procedure Recordings Task Window). If the power is off for more than 1 minute, a non-preset recording automatically reverts to the default waves and recorder for Mode A. However, the speed (25 mm/s), waves non-overlapped, and continuous runtime do not change.

Note—If a realtime recording is requested during an alarm recording, it is produced when the alarm recording is finished. If a realtime recording is in progress, and another parameter goes into alarm, an alarm message appears on the realtime recording. When the realtime recording is finished, the alarm recording is produced.

If you want your ECG recording to be calibrated, you must select the size of the ECG wave, depending on your recorder sector size. To select the size, refer to the “Switching between EASI and Standard leads” on page 14-22. The calibrated ECG wave recording will be 10mm/mV.

Realtime: If a delayed recording, or any recording from another monitor, is in progress at the central station at the time of your request, you are alerted with a message on the monitor display, and the request is queued.

When the recorder is free, the recording runs. It consists of current waveforms (not the waveforms from the time the recording was requested).

If the recorder is busy with a realtime recording and you request another realtime recording from the same monitor, the current recording stops and the newly requested recording starts.

Delayed: If the recorder is busy at the time of your request, the waveform data and the request are queued.

When the recorder is free, the recording runs. The recording shows the waveforms from the time the recording was requested.

Recording Functions

Recording Functions 6-23

Page 195
Image 195
Philips V24CT, V26CT manual Making Calibrated, Recordings If the Recorder is Busy