Philips V26CT, V24CT manual Blood Analysis, Data Transfer, Scenarios

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 308
Image 308

Data Transfer

Types of Transfer

Blood Analysis

• Transferring data from a source monitor to a destination

Data Transfer

monitor where:

Scenarios

 

The source monitor is:

 

— CMS Release G or later or

 

— V24 and V26 Release A.0 or later

The destination monitor is:

 

— CMS Release E or later or

 

— V24 and V26 Release A.0 or later or

 

— V24 Release E

 

1.

Plug the Data Transfer Module into the rack of the source

 

monitor.

 

2.

Press the

 

 

hardkey on the module and transfer all

Module

 

data to the module.

 

3.

Remove the module from the source monitor and plug it into

 

the destination monitor.

4.

Press the

 

 

hardkey on the module and transfer the

Monitor

 

data to the monitor.

 

Note—Not all blood analysis results will be presented if the destination monitor is a CMS Release E or F or a V24 Release E.