The Rear

Panel of the

M1109A

External

Alarm

Device

Controls and

Connectors of

the M1109A

External Alarm

Device

Controls and Connectors

InstallationMonitor

SafetyPatientand

 

 

M1109A External Alarm Device

 

 

4

1

 

2

3

1.Standby Switch

2.Human Interface Link Connector Out; this is a Philips HIL connector used to output the information from the handheld keypad to the Utility CPU function card in the computer module.

3.Human Interface Link Connector In; this is a Philips HIL connector used to connect the handheld keypad.

4.Power on/off cable; this connects to the XGA Display Controller Board C’DSPC’-XGA in the M1046B Computer Module.

Notes—

The connector cover must be put back on after all cables have been connected

The External Alarm Device should be mounted for maximum visibility of the alarm indicators on the front. The device should not be mounted directly against a wall in order to ensure that the alarm

Monitor Installation and Patient Safety 11-29