Alarm Recording

Channel 4

Respiration

A yellow CO2 alarm would produce a recording ordered as:

Channel 1

ECG

Channel 2

Pressure 1

Channel 3

Pressure 2

Channel 4

CO2 alarm

If a red P3 alarm occurred, the recording order would then become:

Channel 1

ECG

Channel 2

Pressure 1

Channel 3

P3 alarm

Channel 4

CO2 alarm

For alarm recordings with oxyCRG (to document neonatal events without the Neonatal Event capability), the alarm recording items should be set to the following:

AlRecType

oxyCRG

All AlarmRec

Off

 

Change AlarmRec to red for ECG, Resp and

 

SpO2.

In the event of a Brady, Apnea and/or Desat alarm the recorder will document an eight minute oxyCRG recording (six minutes preceding the event and two minutes following the event).