Viewing Neonatal Events

Event Review

19 Stored Events: A(D): 4(2) B(D): 3(0) AB(D): 1(1) D: 9 M: 2

Example of an Event Summary Line

The above example of an event summary line, conveys the following information:

A total of 19 neonatal events are stored.

Four Apnea events occurred, of which two resulted in a Desaturation.

Three isolated Bradycardia events occurred, none of them resulted in a Desaturation.

One associated Apnea and Bradycardia occurred which also resulted in a Desaturation.

Nine isolated Desaturations occurred.

Two events were manually captured and stored.

Event Storage If the number of stored events reaches a total of 441, the following status message is displayed in the Event Review task window:

“Event Storage>90% full”

If the number of stored events exceeds a total of 482, the oldest events are overwritten and the following status message is displayed in the Event Review task window:

“Event Storage overflow. First events overwritten”

Neonatal Event

Review

 

 

1.22 for V24 and V26 Patient Monitors

2.24 for V24 and V26 Patient Monitors

Neonatal Event Review 9-11