Setting up the Monitor (V24 and V26 only)

2. If you are using the V24CT and V26CT and the batteries are not already in place, insert 2 Philips 40488A 12 Volt 2.3 Amp-hour sealed lead-acid batteries into the spring-loaded battery door in the back of the monitor. The batteries should be inserted with the grey Philips battery label facing upwards and the contacts pointing into the monitor. The batteries will snap or click into place when properly

inserted. StartedGetting

00000A

C aution

Philips Medical Systems recommends you fully charge the batteries before using the V24CT or V26CT for the first time. Plug the monitor into an AC power supply (line power) for an initial charging cycle. When the monitor is plugged in and the power is off, it takes approximately four hours for the batteries to reach a charge of 90% capacity and nine hours for the batteries to be fully charged. If the monitor is plugged in and the power and external battery charger is on, it takes approximately 16 hours to charge the batteries.

3.If using an AC power supply, plug the power cord into the power source.