Philips V26CT, V24CT manual V26 Connectors

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 350
Image 350

Monitor Installation

and Patient Safety

 

 

Controls and Connectors

The V24 and The monitor has several connectors which are shown below.

V26

Connectors

DEFIB

CONNECTOR

AC LINE INPUT

RACK

 

CONNECTOR

RS232 CONNECTOR (Rear Housing with Serial Prefix <3608)

PROTECTIVE

 

EARTH GROUND

EQUIPOTENTIAL GROUNDING

RS232 CONNECTOR

(Rear Housing with PHILIPS MONITORING NETWORK (SDN)

Serial Prefix >=3608)

SINGLE CONNECTOR

Do not apply a voltage of more than +12V to any of the pins on the defibrillator sync, parameter module rack, Philips monitoring network (SDN) or equipotential ground connectors.

The secondary ground complies with IEC 601-1-1 as shown on page 10-4.

The connectors on the monitor are:

1.AC Power Connector; this connector is used to connect the monitor to an AC power source.

Note—If you have a Philips M1026A Anesthetic Gas Module (AGM), do not daisy-chain the AC power through the AGM. This configuration may degrade RFI performance.

2.Secondary Ground Screw; this is a posidrive screw to connect the monitor chassis to earth ground per IEC 601-1-1 standard.

3.Equipotential Ground Post, this connects the monitor to the equipotential ground system.

11-34 Monitor Installation and Patient Safety