The CMS and V24 and V26 Patient Monitors

Using an ITE Display

Using an ITE Display

The M1167A/77A CMS Patient Monitoring System provides means for customers to use ITE displays. There are several restrictions associated with this option. Since the display is used as part of a medical device within the patient vicinity, there are additional requirements to be fulfilled in order to be in compliance with the European Council Directive 93/42/EEC (Medical Device Directive) and with FDA recognized consensus standards:

 

Safety

 

EMC

 

Performance Requirements

Safety

1.

The display must either comply with the requirements of the:

 

 

EN60601-1 +A1 +A2 {IEC601-1 +A1 +A2}, or

 

 

it must conform to the requirements of an IEC XXX electrical

 

 

safety standard and fulfill specific requirements according to the

 

 

EN60601-1-1 +A1 {IEC601-1-1 + A1} to provide the same level of

 

 

safety as EN60601-1 {IEC601-1}:

 

 

– The display must be constructed or protected so that spillage

 

 

of liquids does not wet parts of the display which may cause a

 

 

safety hazard (IEC601-1; 44.3).

 

 

– The ITE display must be used with an isolation transformer

 

 

(e.g. Philips M1389A), if the enclosure leakage current

 

 

exceeds the requirements of EN60601-1/IEC601-1 (normal

 

 

and single fault condition). The power cable must be secured

 

 

so that the transformer cannot be disconnected without the

 

 

use of a tool.

 

2.

A test of the Enclosure Leakage Current (normal and single fault

 

 

condition) of the complete system, including the display, has to be

 

 

performed and documented.

1-38 The CMS and V24 and V26 Patient Monitors

Page 60
Image 60
Philips V26CT, V24CT manual Using an ITE Display, Safety, Performance Requirements