Alarm Functions

Alarm Display

 

Alarm

Red alarms - Highest priority alarms, displayed in the top right

Priorities

corner of the display screen. These are marked with three

 

asterisks (***) next to the alarm message, for example,

 

***ASYSTOLE.

 

Yellow alarms - displayed in the top center of the display screen.

 

There are two types of yellow alarms; continuous yellow alarms

 

and short yellow alarms. The continuous yellow alarms are

 

associated with preset alarm limits. The short yellow alarms are

 

active from the arrhythmia computer.

 

Both types of yellow alarms are marked with two asterisks (**)

 

next to the alarm message, for example, **HR 56<60, but the alarm

 

tone is shorter with the alarms from the arrhythmia computer. For

 

both the short and continuous yellow alarm, the yellow lamp is lit.

 

INOPs - displayed in the top left corner of the display screen. The

 

INOPs also have priority levels, they are as follows:

 

– Equipment Related INOPs - These are caused by setup or

 

hardware faults, for example, “ABP NO TRANSDUCER”. You

 

will hear an INOP tone. When an equipment-related INOP can

 

have a direct effect on the patient, for example “NBP CUFF

 

OVERPRESS” of the noninvasive pressure cuff, it is

 

categorized as a severe INOP. When a severe INOP occurs

 

the system alarm capability is switched ON by the system, if

 

not already on. You will hear an INOP tone. System alarms

 

cannot be suspended until the cause of the severe INOP has

 

been eliminated.

 

– Patient Related INOPs - These are usually caused by the

 

system being unable to process signals properly, for example,

 

“Noisy ECG”. You will not hear an INOP tone with this level

 

of INOP.

5-8 Alarm Functions