Controls and Connectors

4.Philips monitoring network connector (SDN Interface); this is a proprietary Philips system connector, used to input/output information to a Philips monitoring network.

Warning

Connecting the Philips monitoring network (SDN) cable when the product is powered on is not supported. Error codes and Philips monitoring network (SDN) interface lock-up may occur. Power cycling the product will recover the product. No permanent damage will result. To prevent unintentional disruption in monitoring, be sure the SDN interface cable is properly secured at both ends when connecting to the Philips monitoring network (SDN).

Monitor Installation

and Patient Safety

 

 

5.Parameter Module Rack Connector; this connects the parameter module rack to the monitor.

6.Defibrillator sync connector; sends an ECG signal out to a defibrillator or other external device or receives a marker pulse from a defibrillator or other external device. The marker pulse is processed with the ECG signal and displayed on the monitor.

Note—To filter radiated emissions, a core shielding bead (Philips P/N 9170-1521) must be attached to the defibrillator synchronization cable when a defibrillator is connected to the monitor. This filter has no impact on the function of the monitor. To attach the core shielding bead:

a.Open the core shielding bead (filter).

b.Place the bead around the defib sync cable as close to the jack as possible.

c.Close the bead over the cable and snap shut.

d.Plug the cable into the monitor's defib sync port. The bead should be at the monitor end of the cable.

Monitor Installation and Patient Safety 11-35