Philips V26CT, V24CT manual General cleaning of the System, Soaps, Ammonias, Alcohol

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 368
Image 368

Maintenance

General cleaning of the System

General cleaning of the System

The patient monitor should be kept free of dust and dirt.

Exterior cleaning of the case and screen is recommended. Clean it with a lint-free cloth, moistened with either warm water (40°C/104°F. max) and soap, a diluted non-caustic detergent or one of the approved cleaning agents listed below:

Soaps

mild soaps

Tensides (dishwasher

Edisonite Schnellreiniger, Alconox

detergents)

 

Ammonias

Dilution of Ammonia <3%, Window

 

cleaner

Alcohol

Ethanol 70%, Isopropanol 70%, Window

 

cleaner

C aution

To avoid damage to the product, observe the following general precautions for cleaning. You should only deviate when this is explicitly described in the cleaning instruction for the individual transducer or accessory.

1.Do not use strong solvents such as acetone or trichloroethylene.

2.Always dilute according to the manufacturers instructions, or use lowest possible concentration.

3.Never use abrasive material (such as steel wool or silver polish).

4.Do not allow liquid to enter into the product.

5.Never submerge any part of the system.

6.Do not pour liquid onto the system during cleaning.

13-2 Maintenance

Page 368
Image 368