Setting up your Monitor

Changing the Patient Category

Recommendation

Choose the patient category according to the QRS morphology, rather than the patient's age. Use Adult if the duration of the QRS base is 50 to 70 ms: use Neonate/Pediatric if the duration is less than 50 ms.

Weak signal The signal may be weak due to the patient's constitution, the electrode placement or other factors. A 1 mV calibration bar at the left side of the ECG channel enables you to estimate the quality of the signal.

Recommendation

Use the strongest signal on channel 1 and compare the QRS amplitude with the 1 mV bar; the height should be at least half of that size. If not, change the lead or try a different electrode placement.

Paced Patients Paced pulse rejection is available for patient categories.

Recommendations

Derive the cardiac rate from pulse.

Look for a lead on channel 1, where the pace pulse amplitude is zero to avoid paced pulses being counted as beats.

Whenever acceptable use the Adult category.

3-50 Setting up your Monitor